Abstract

In this study, a simple and reliable stability-indicating RP-HPLC method was developed and validated for the analysis of Nystatin in the pharmaceuticals. The chromatographic separation was performed in the isocratic mode on an Ion Pac column; Arcus EP‑C18; 5μm, 4.6×250 mm, 30 °C) using a mobile phase consisting of ammonium acetate 0.05 M buffer/ Methanol mixture (30:70) and a flow-rate of 1.0 mL/min with UV detection at 305 nm. The flow rate was set at 1.0 mL/min. The HPLC analysis method was validated in terms of linearity, precision, accuracy, specificity, and sensitivity, according to International Conference on Harmonization (ICH) guidelines. The results indicated that the retention time was 8 min, and no interferences were observed from the formulation excipients and stress degradation products. The specificity, linearity, precision, accuracy, LOD, and LOQ of the method were validated. The method was linear over the range of 5–500 μg/mL with an acceptable correlation coefficient (R2 = 0.9996). The method’s limit of detection (LOD) and quantification (LOQ) were 0.01 and 0.025 μg/mL, respectively. The results indicate that this validated method can be used as an alternative method for the assay of nystatin. This validated HPLC method could be used for routine analysis, quality control, and the stability of analysis of Nystatin formulations.

Full text article

Generated from XML file

Authors

Dawood CH. Al-Bahadily
Rasool Chaloob
Kulood H. Oudah
H. N. K. AL-Salman
hsennaserh@yahoo.com (Primary Contact)
Falah Hassan Shari
Hussein H. Hussein
Dawood CH. Al-Bahadily, Rasool Chaloob, Kulood H. Oudah, H. N. K. AL-Salman, Falah Hassan Shari, & Hussein H. Hussein. (2019). Fasten, simple, and specific stability of the avant-garde RP-HPLC method for estimation and validation of nystatin in pharmaceutical formulations. International Journal of Research in Pharmaceutical Sciences, 10(4), 3717–3727. Retrieved from https://ijrps.com/home/article/view/3301

Article Details

No Related Submission Found